DOI | Resolve DOI: https://doi.org/10.1109/TIM.2020.3045797 |
---|
Author | Search for: Djokic, Branislav1ORCID identifier: https://orcid.org/0000-0001-7546-9207; Search for: Parks, Harold1ORCID identifier: https://orcid.org/0000-0002-8144-5460 |
---|
Affiliation | - National Research Council of Canada. Metrology Research Centre
|
---|
Format | Text, Article |
---|
Subject | calibration; harmonics; measurement; measurement uncertainty; power quality; power quality analyzer; harmonic analysis; power system harmonics; frequency measurement; power measurement; voltage measurement |
---|
Abstract | The paper describes the recent development of a system for calibration of electrical instruments with harmonics measurement capability, such as revenue meters and power analyzers, under nonsinusoidal test conditions at National Research Council (NRC) Canada. The improvements implemented with respect to an earlier NRC calibration system are aimed at increasing the calibration accuracy in a wider frequency range. |
---|
Publication date | 2020-12-25 |
---|
Publisher | IEEE |
---|
In | |
---|
Language | English |
---|
Peer reviewed | Yes |
---|
Export citation | Export as RIS |
---|
Report a correction | Report a correction (opens in a new tab) |
---|
Record identifier | 1f6e3f12-19a1-41bf-a3f6-8fdcf36bd5dc |
---|
Record created | 2023-01-04 |
---|
Record modified | 2023-01-04 |
---|