| DOI | Resolve DOI: https://doi.org/10.1016/S0584-8547(01)00263-4 |
|---|
| Author | Search for: Willie, S. N.1; Search for: Sturgeon, R.1 |
|---|
| Affiliation | - National Research Council of Canada. NRC Institute for National Measurement Standards
|
|---|
| Format | Text, Article |
|---|
| Subject | inductively coupled plasma; time-of-flight mass spectrometry; trace elements; rare earth elements; flow injection; on-line separation and preconcentration; seawater |
|---|
| Abstract | A method is presented for the quantitation of trace elements and rare earth elements (REE) in seawater by inductively coupled plasma orthogonal acceleration time-of-flight mass spectrometry. On-line separation and preconcentration was achieved using a timed flow-injection (FI) system incorporating a column containing iminodiacetate-based resin. The procedure was validated for selected transition elements by analysis of certified seawater reference materials NASS-5 and CASS-4. Recoveries of the rare earth elements (REE) from seawater were between 96 and 101%. Limits of detection (3s) were generally in the 20–50-pg l−1 range with nominal precision of 5–6% R.S.D. (n=5) when processing 50-ml sample aliquots of seawater. Sample throughput was 5 h−1 for the REE determinations and 15 h−1 for the transition elements of interest. |
|---|
| Publication date | 2001-09-30 |
|---|
| In | |
|---|
| Language | English |
|---|
| NRC number | NRC-INMS-160 |
|---|
| NPARC number | 5763129 |
|---|
| Export citation | Export as RIS |
|---|
| Report a correction | Report a correction (opens in a new tab) |
|---|
| Record identifier | 2f68562f-e20b-4ea8-8a20-f8b67c85f131 |
|---|
| Record created | 2009-03-29 |
|---|
| Record modified | 2020-03-27 |
|---|