| DOI | Resolve DOI: https://doi.org/10.1016/j.apradiso.2015.11.068 |
|---|
| Author | Search for: Galea, R.1; Search for: Gameil, K. |
|---|
| Affiliation | - National Research Council Canada. Measurement Science and Standards
|
|---|
| Format | Text, Article |
|---|
| Conference | 20th International Conference on Radionuclide Metrology and its Applications, June 8-11 2015, Vienna, Austria |
|---|
| Subject | Quality assurance; Radionuclide ionization chambers; Hospital radionuclide calibrators |
|---|
| Abstract | From 1986 to 2000, a fraction of the Canadian nuclear medicine community participated in a service offered by the National Research Council (NRC) of Canada to check the accuracy of administered doses of radiopharmaceuticals. The NRC renewed this service in recent years with a revalidation and reviving of the Secondary Standard Ionizing Radiation Chamber System (SSIRCS). The NRC conducted mock services for ⁹⁹ mTc, in varying geometries (syringes and serum vials), on two NRC commercial radionuclide calibrators, and at a nuclear medicine department in Canada. These tests showed the measured doses to be within 10% of the prepared standard but in some cases were sufficiently different from unity to warrant specific geometry factors to be derived. |
|---|
| Publication date | 2015-11-25 |
|---|
| Publisher | Elsevier |
|---|
| In | |
|---|
| Language | English |
|---|
| Peer reviewed | Yes |
|---|
| Identifier | S0969804315302931 |
|---|
| NPARC number | 23000073 |
|---|
| Export citation | Export as RIS |
|---|
| Report a correction | Report a correction (opens in a new tab) |
|---|
| Record identifier | 55b584c0-d7d7-4e94-ade1-9f97574baf41 |
|---|
| Record created | 2016-06-01 |
|---|
| Record modified | 2020-04-22 |
|---|