Alternative title | Determination of Wair in high-energy electron beams using graphite detectors |
---|
Download | - View accepted manuscript: Determination of Wₐᵢᵣ in high‐energy electron beams using graphite detectors (PDF, 1.8 MiB)
|
---|
DOI | Resolve DOI: https://doi.org/10.1002/mp.13772 |
---|
Author | Search for: Bourgouin, Alexandra; Search for: Cojocaru, Claudiu1; Search for: Ross, Carl1; Search for: McEwen, Malcolm1 |
---|
Affiliation | - National Research Council of Canada. Metrology Research Centre
|
---|
Format | Text, Article |
---|
Physical description | 14 p. |
---|
Subject | average energy to create an ion pair in air; EGSnrc; graphie calorimeter; ionization chamber; megavoltage electrons beam; thermal simulation |
---|
Abstract | ICRU Report 90 on Key Data for Ionizing‐Radiation Dosimetry: Measurement Standards and Applications (2014) has reaffirmed the recommended value of the mean energy required to create an ion pair in air, Wₐᵢᵣ, to be 33.97(12) eV. The report also indicates that this “constant” of radiation dosimetry is energy independent above 10 keV, since there is no theoretical or experimental evidence to the contrary. The goal of this investigation is to obtain additional experimental determinations of Wₐᵢᵣ in high energy beams and thus to verify the suggested energy independence. |
---|
Publication date | 2019-09-28 |
---|
Publisher | Wiley on behalf of the American Association of Physicists in Medicine |
---|
In | |
---|
Language | English |
---|
Peer reviewed | Yes |
---|
Export citation | Export as RIS |
---|
Report a correction | Report a correction (opens in a new tab) |
---|
Record identifier | 5a3e975b-b00e-4aaf-a2e4-57c14f660d93 |
---|
Record created | 2019-10-30 |
---|
Record modified | 2020-08-31 |
---|