Download | - View accepted manuscript: Cobalt–carbon eutectic fixed point for contact thermometry (PDF, 3.0 MiB)
|
---|
DOI | Resolve DOI: https://doi.org/10.1007/s10765-010-0907-5 |
---|
Author | Search for: Todd, A. D. W.1; Search for: Gotoh, M.1; Search for: Woods, D. J.1; Search for: Hill, K. D.1 |
---|
Affiliation | - National Research Council of Canada. NRC Institute for National Measurement Standards
|
---|
Format | Text, Article |
---|
Subject | Eutectics; High-temperature fixed points; Pd–C; Pt/Pd Thermocouples |
---|
Abstract | Two Co–C eutectic fixed points were constructed for thermocouple calibration. The eutectic fixed points were measured with a Pt/Pd thermocouple calibrated at the freezing temperatures of Sn, Zn, Al, Ag, and Au. A temperature of (1323.99 ± 0.52)◦C(k = 2) was determined via thismethod. The cell design allowed filling to be accomplished in a single step. Each cellwas held above 1300 ◦C for at least 42 h and was subjected to at least 20 melt/freeze cycles with no mechanical failure occurring.abstract |
---|
Publication date | 2010-12-29 |
---|
In | |
---|
Language | English |
---|
Peer reviewed | Yes |
---|
NPARC number | 17320716 |
---|
Export citation | Export as RIS |
---|
Report a correction | Report a correction (opens in a new tab) |
---|
Record identifier | 86860a58-69d2-4df0-a7b9-a8bf77d3c1ec |
---|
Record created | 2011-04-03 |
---|
Record modified | 2020-04-17 |
---|