| Download | - View author's version: Calibration of electrical instruments under non-sinusoidal conditions at NRC Canada (PDF, 464 KiB)
|
|---|
| DOI | Resolve DOI: https://doi.org/10.1109/CPEM49742.2020.9191765 |
|---|
| Author | Search for: Djokic, Branislav1; Search for: Parks, Harold1 |
|---|
| Affiliation | - National Research Council Canada. Metrology Research Centre
|
|---|
| Format | Text, Article |
|---|
| Conference | 2020 Conference on Precision Electromagnetic Measurements (CPEM 2020), August 24-28, 2020, Denver (Aurora), CO, USA |
|---|
| Subject | calibration; harmonic analysis; power system harmonics; power measurement; standards; power quality; frequency measurement |
|---|
| Abstract | The paper describes a recent development of a system for calibration of electrical instruments with harmonics measurement capability, such as revenue meters and power analyzers, under non-sinusoidal test conditions at National Research Council (NRC) Canada. The improvements implemented with respect to an earlier NRC calibration system are aimed at increasing the calibration accuracy and frequency range. |
|---|
| Publication date | 2020-08-24 |
|---|
| Publisher | IEEE |
|---|
| In | |
|---|
| Language | English |
|---|
| Peer reviewed | Yes |
|---|
| Export citation | Export as RIS |
|---|
| Report a correction | Report a correction (opens in a new tab) |
|---|
| Record identifier | a12d0fa6-f27c-410c-a17d-c8d5d07c01b0 |
|---|
| Record created | 2020-09-15 |
|---|
| Record modified | 2020-09-15 |
|---|