DOI | Resolve DOI: https://doi.org/10.1118/1.3314072 |
---|
Author | Search for: Tessier, Frédéric1; Search for: Hooten, Brian D.; Search for: McEwen, Malcolm R.1 |
---|
Affiliation | - National Research Council of Canada. NRC Institute for National Measurement Standards
|
---|
Format | Text, Article |
---|
Subject | biomedical equipment; dosimetry; ionisation chambers; effective point of measurement; ionization chambers |
---|
Abstract | Purpose: Verify experimentally the theoretical prediction of F. Tessier and I. Kawrakow [Med. Phys. 37, 96–107 (2010)] that it is possible to design a thimble ionization chamber with no shift in its effective point of measurement (EPOM), i.e., a chamber that provides a measure of the dose to the medium at the location of its central axis. |
---|
Publication date | 2010-02-18 |
---|
In | |
---|
Language | English |
---|
Peer reviewed | Yes |
---|
NPARC number | 16378394 |
---|
Export citation | Export as RIS |
---|
Report a correction | Report a correction (opens in a new tab) |
---|
Record identifier | ca3e8b35-b2a2-4fe5-a1a4-42235415b74e |
---|
Record created | 2010-11-15 |
---|
Record modified | 2020-04-17 |
---|