DOI | Resolve DOI: https://doi.org/10.1109/TIM.2016.2631739 |
---|
Author | Search for: Djokic, Branislav V.1; Search for: Parks, Harold1; Search for: Wise, Nicholas1; Search for: Naumovic-Vukovic, Dragana; Search for: Skundric, Slobodan P.; Search for: Zigic, Aleksandar D.; Search for: Poluzanski, Vladimir |
---|
Affiliation | - National Research Council of Canada. Measurement Science and Standards
|
---|
Format | Text, Article |
---|
Subject | AC current measurement; calibration; current comparator; current transformer; measurement uncertainty |
---|
Abstract | A comparison of two systems based on different methods for calibration of current transformers at several different transformation ratios, burdens, and power frequencies of 60 Hz and 50 Hz is performed at the National Research Council of Canada. The obtained discrepancies in ratio and phase errors are within 5 μA/A and μrad, respectively, for small test currents and less for larger test currents. |
---|
Publication date | 2016-12-26 |
---|
Publisher | IEEE |
---|
In | |
---|
Language | English |
---|
NPARC number | 23002557 |
---|
Export citation | Export as RIS |
---|
Report a correction | Report a correction (opens in a new tab) |
---|
Record identifier | ea15a063-d200-4199-853c-6d49e2e4434b |
---|
Record created | 2017-11-30 |
---|
Record modified | 2020-03-16 |
---|