Téléchargement | - Voir la version finale : Calibration model averaging in chemical analysis: a case study for the method of standard additions (PDF, 780 Kio)
- Voir les données supplémentaires : Calibration model averaging in chemical analysis: a case study for the method of standard additions (PDF, 1.7 Mio)
|
---|
DOI | Trouver le DOI : https://doi.org/10.1088/1681-7575/accd75 |
---|
Auteur | Rechercher : Pagliano, Enea1Identifiant ORCID : https://orcid.org/0000-0001-9755-8737; Rechercher : Meija, Juris1Identifiant ORCID : https://orcid.org/0000-0002-3349-5535 |
---|
Affiliation | - Conseil national de recherches du Canada. Centre de recherche en métrologie
|
---|
Format | Texte, Article |
---|
Sujet | method of standard additions; uncertainty evaluation; model averaging |
---|
Résumé | In this study we demonstrate that the method of standard additions can provide unreliable results in chemical analysis when the linear calibration model is forced to fit the data. A well designed standard addition experiment can still yield results biased by 10% when the analyst relies only on the linear model. Recently, the Joint Committee for Guides in Metrology (GUM-6) has emphasized the need to address the uncertainty inherent to the choice of measurement models and here we show how model averaging can provide a practical way to account for model uncertainty in the method of standard additions. |
---|
Date de publication | 2023-05-11 |
---|
Maison d’édition | IOP Publishing |
---|
Licence | |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | 035380be-0529-44b8-a20c-e7ceb31f846b |
---|
Enregistrement créé | 2024-01-25 |
---|
Enregistrement modifié | 2024-01-25 |
---|