DOI | Trouver le DOI : https://doi.org/10.1109/CPEM.2018.8501152 |
---|
Auteur | Rechercher : Granger, G.1, 2; Rechercher : Budovsky, I. |
---|
Affiliation | - Conseil national de recherches du Canada. Science des mesures et étalons
- Conseil national de recherches du Canada. Centre de recherche en métrologie
|
---|
Format | Texte, Article |
---|
Conférence | 2018 Conference on Precision Electromagnetic Measurements (CPEM 2018), 8-13 July 2018, Paris, France |
---|
Sujet | AC-DC difference; Josephson Arbitrary Waveform Synthesizer; Josephson Voltage Standard; pulsedriven; thermal transfer standard; transconductance amplifier |
---|
Résumé | A transconductance amplifier (TCA) with a negative admittance has been used in parallel with a commercial thermal transfer standard (TTS) to unload the voltage source, a pulse-driven AC Josephson Voltage Standard (also known as Josephson Arbitrary Waveform Synthesizer or JAWS). AC-DC difference measurements at 1 V and at frequencies from 10 Hz to 10 kHz show excellent agreement with the values obtained by a conventional method. Expanded uncertainties between 0.28 µV/V and 0.39 µV/V (k=2), depending on the frequency, have been achieved. |
---|
Date de publication | 2018-10-22 |
---|
Maison d’édition | IEEE |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | 1e55c830-5639-40c1-8033-aba3f190a954 |
---|
Enregistrement créé | 2019-03-22 |
---|
Enregistrement modifié | 2020-03-16 |
---|