DOI | Trouver le DOI : https://doi.org/10.1109/CPEM.2016.7540556 |
---|
Auteur | Rechercher : Jung, Jae Kap; Rechercher : Lee, Young Seob; Rechercher : Kim, Kyu-Tae; Rechercher : So, Eddy1 |
---|
Affiliation | - Conseil national de recherches du Canada. Science des mesures et étalons
|
---|
Format | Texte, Article |
---|
Conférence | 2016 Conference on Precision Electromagnetic Measurements (CPEM 2016), July 10-15, 2016, Ottawa, ON, Canada |
---|
Sujet | calibration; Capacitor; dissipation factor; capacitance bridge; measurement uncertainty |
---|
Résumé | An absolute method for determining the dissipation factor (DF) of a capacitor connected to a series resistor without the use of a reference capacitor with a known DF is described. The principle of the absolute measurement is based on the “elimination” of DF value of a reference capacitor by subtracting reading values from two successive measurements. This method was applied to the calibration of DF within a range of 1 × 10-5 to 1 × 10-2 at voltages up to 1 kV The verification of this method was confirmed by comparing the obtained values of DF with that measured by a known calibrated multi-frequency capacitance meter. |
---|
Date de publication | 2016-07 |
---|
Maison d’édition | IEEE |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Numéro NPARC | 23002540 |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | 26383386-3700-4267-b468-ebc0c22b6275 |
---|
Enregistrement créé | 2017-11-27 |
---|
Enregistrement modifié | 2020-03-16 |
---|