DOI | Trouver le DOI : https://doi.org/10.1088/0026-1394/48/3/010 |
---|
Auteur | Rechercher : Woolliams, E.R.; Rechercher : Baribeau, R.1; Rechercher : Bialek, A.; Rechercher : Cox, M.G. |
---|
Affiliation | - Conseil national de recherches du Canada. Institut des étalons nationaux de mesure du CNRC
|
---|
Format | Texte, Article |
---|
Sujet | Band pass; Band-pass function; Finite bandwidth; Measurement Noise; Practical method; Radiometric measurements; Bandwidth; Spectrometry; Spectrometers |
---|
Résumé | The bandpass of spectrometers can cause appreciable errors when making radiometric measurements. This paper describes a practical method for correcting a set of equispaced measured values provided by a spectrometer with a finite bandwidth, an arbitrary bandpass function and at an arbitrary wavelength step. The paper reviews the limits of the approach for real spectra in the presence of measurement noise and suggests ways of reducing the effect of noise. © 2011 BIPM & IOP Publishing Ltd. |
---|
Date de publication | 2011 |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Numéro NPARC | 21271078 |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | 269433a7-9391-48a9-bb91-05a0caef5f08 |
---|
Enregistrement créé | 2014-03-24 |
---|
Enregistrement modifié | 2020-04-21 |
---|