DOI | Trouver le DOI : https://doi.org/10.1109/NSSMIC.2011.6154535 |
---|
Auteur | Rechercher : Boyle, P. J.; Rechercher : Hanna, D. S.; Rechercher : Macleod, A. M. L.; Rechercher : Sinclair, L. E.; Rechercher : Seywerd, H. C. J.; Rechercher : Saull, P. R. B.1 |
---|
Affiliation | - Conseil national de recherches du Canada
|
---|
Format | Texte, Article |
---|
Conférence | 2011 IEEE Nuclear Science Symposium and Medical Imaging Conference (2011 NSS/MIC), 2011-10-23 - 2011-10-29, Valencia, Spain |
---|
Sujet | gamma-ray detection; nuclear electronics; solid scintillation detectors |
---|
Résumé | In this work we present an optimization of the energy and position resolutions for use in gamma-ray interactions in a long bar of NaI(TI). This work focuses on tuning the effective optical attenuation coefficient to approximately the same length of the bar. This is achieved by altering the surface of the scintillator in a systematic manner. We present data obtained in the laboratory from measurements of eighteen bars of NaI(TI) that have been manufactured with various treatments of the surface. This process results in an improvement in position resolution by a factor of four over a polished bar with only a corresponding 20% decrease in energy resolution. |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Numéro NPARC | 21269139 |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | 3dc97f0e-0478-4c32-b228-831e231bec56 |
---|
Enregistrement créé | 2013-12-09 |
---|
Enregistrement modifié | 2020-04-22 |
---|