DOI | Trouver le DOI : https://doi.org/10.1063/1.4819607 |
---|
Auteur | Rechercher : Gotoh, M.1 |
---|
Affiliation | - Conseil national de recherches du Canada. Science des mesures et étalons
|
---|
Format | Texte, Article |
---|
Conférence | Ninth International Temperature Symposium, 19–23 March 2012, Los Angeles, California, USA |
---|
Sujet | tungsten/rhenium thermocouple; Co-C eutectic point; Pd-C eutectic point |
---|
Résumé | Four Type A thermocouples and two Type C thermocouples were calibrated at the Au fixed point and Co-C and Pd-C eutectic fixed points. The thermocouples were exposed to 1330 °C for a total of 100 hours. The maximum drift due to the exposure was found to be 4.8 °C. The fixed-point calibration EMF of these thermocouples deviated by less than 0.86% from the temperature specified by the standards ASTM E230-2003 for Type C and GOSTR 8.585-2001 for Type A. The length of one of Type A thermocouples A52 is longer than the others by 150mm. Making use of this provision it was possible to place annealed part of A52 to the temperature gradient part of calibration arrangement every time. Therefore observed aging effect was as low as 0.5 °C compared to the other thermocouples. |
---|
Date de publication | 2013-09-11 |
---|
Dans | |
---|
Série | |
---|
Note | M. Gotoh was a visiting scientist at the time of publication |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Numéro du CNRC | NRC-MSS-1569481257 |
---|
Numéro NPARC | 21268886 |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | 4b9ddded-9514-4432-94ad-a51d6d7d8f40 |
---|
Enregistrement créé | 2013-11-22 |
---|
Enregistrement modifié | 2020-04-22 |
---|