DOI | Trouver le DOI : https://doi.org/10.1109/EFTF/IFCS52194.2021.9604288 |
---|
Auteur | Rechercher : Dube, Pierre1; Rechercher : Kato, Kosuke1; Rechercher : Bernard, John1; Rechercher : Jian, Bin1 |
---|
Affiliation | - Conseil national de recherches du Canada. Centre de recherche en métrologie
|
---|
Format | Texte, Article |
---|
Conférence | 2021 Joint Conference of the European Frequency and Time Forum and IEEE International Frequency Control Symposium (EFTF/IFCS), July 7-17, 2021, Gainesville, FL, USA |
---|
Sujet | ion optical clock; ion trap design; blackbody radiation shift; Particle beam optics; Time-frequency analysis; Uncertainty; Optical polarization; Optical design; Ions; Optical variables control |
---|
Résumé | We are developing a transportable ⁸⁸Sr⁺ ion clock for frequency comparisons with optical clocks in remote laboratories. Accuracy limitations found in the current NRC laboratory system, most notably the uncertainties related to the blackbody radiation shift and the collisional frequency shifts, should be significantly reduced in the new system. The expected total systematic fractional frequency uncertainty of the device approaches the 10⁻¹⁸ level. |
---|
Date de publication | 2021-07-07 |
---|
Maison d’édition | IEEE |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | 6455f451-882f-460f-8872-7e1a80ce1413 |
---|
Enregistrement créé | 2023-04-17 |
---|
Enregistrement modifié | 2023-04-17 |
---|