DOI | Trouver le DOI : https://doi.org/10.1098/rsta.2015.0044 |
---|
Auteur | Rechercher : Woolliams, E. R.; Rechercher : Anhalt, K.; Rechercher : Ballico, M.; Rechercher : Bloembergen, P.; Rechercher : Bourson, F.; Rechercher : Briaudeau, S.; Rechercher : Campos, J.; Rechercher : Cox, M. G.; Rechercher : Del Campo, D.; Rechercher : Dong, W.; Rechercher : Dury, M. R.; Rechercher : Gavrilov, V.; Rechercher : Grigoryeva, I.; Rechercher : Hernanz, M. L.; Rechercher : Jahan, F.; Rechercher : Khlevnoy, B.; Rechercher : Khromchenko, V.; Rechercher : Lowe, D. H.; Rechercher : Lu, X.; Rechercher : Machin, G.; Rechercher : Mantilla, J. M.; Rechercher : Martin, M. J.; Rechercher : McEvoy, H. C.; Rechercher : Rougié, B.; Rechercher : Sadli, M.; Rechercher : Salim, S. G. R.; Rechercher : Sasajima, N.; Rechercher : Taubert, D. R.; Rechercher : Todd, A. D. W.1; Rechercher : Van den Bossche, R.; Rechercher : van der Ham, E.; Rechercher : Wang, T.; Rechercher : Whittam, A.; Rechercher : Wilthan, B.; Rechercher : Woods, D. J.1; Rechercher : Woodward, J. T.; Rechercher : Yamada, Y.; Rechercher : Yamaguchi, Y.; Rechercher : Yoon, H. W.; Rechercher : Yuan, Z. |
---|
Affiliation | - Conseil national de recherches du Canada. Science des mesures et étalons
|
---|
Format | Texte, Article |
---|
Sujet | High-temperature fixed points; Thermodynamic temperature; Thermometry; Temperature scale; Kelvin; Eutectics |
---|
Résumé | The thermodynamic temperature of the point of inflection of the melting transition of Re-C, Pt-C and Co-C eutectics has been determined to be 2747.84 ± 0.35 K, 2011.43 ± 0.18 K and 1597.39 ± 0.13 K, respectively, and the thermodynamic temperature of the freezing transition of Cu has been determined to be 1357.80 ± 0.08 K, where the ± symbol represents 95% coverage. These results are the best consensus estimates obtained from measurements made using various spectroradiometric primary thermometry techniques by nine different national metrology institutes. The good agreement between the institutes suggests that spectroradiometric thermometry techniques are sufficiently mature (at least in those institutes) to allow the direct realization of thermodynamic temperature above 1234 K (rather than the use of a temperature scale) and that metal-carbon eutectics can be used as high-temperature fixed points for thermodynamic temperature dissemination. The results directly support the developing mise en pratique for the definition of the kelvin to include direct measurement of thermodynamic temperature. |
---|
Date de publication | 2016-02-22 |
---|
Maison d’édition | The Royal Society |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Numéro NPARC | 23000081 |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | 73d31b55-8679-4ea9-b679-4fc585e8b76e |
---|
Enregistrement créé | 2016-06-02 |
---|
Enregistrement modifié | 2020-03-16 |
---|