Téléchargement | - Voir le manuscrit accepté : Correction of systematic errors due to the voltage leads in an AC Josephson Voltage Standard (PDF, 742 Kio)
|
---|
DOI | Trouver le DOI : https://doi.org/10.1109/TIM.2008.2008086 |
---|
Auteur | Rechercher : Filipski, Piotr S.1; Rechercher : Kinard, Joseph R.; Rechercher : Lipe, Thomas E.; Rechercher : Tang, Yi-Hua; Rechercher : Benz, Samuel P. |
---|
Affiliation | - Conseil national de recherches du Canada
|
---|
Format | Texte, Article |
---|
Résumé | The National Institute of Standards and Technology (NIST) has recently reported the first application of a quantum ac Josephson voltage standard for the calibration of thermal transfer standards in the 1- to 10-kHz frequency range. This paper describes preliminary work on extending its frequency calibration range up to 100 kHz by correcting the systematic errors due to voltage leads. A ground loop created by the dc blocks, which is a previously unaccounted source of high-frequency systematic error, has been identified, and its effects are partially mitigated. |
---|
Date de publication | 2009-04 |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Numéro NPARC | 18116653 |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | 848bde35-0377-4db9-9e4e-2c488274e53a |
---|
Enregistrement créé | 2012-03-09 |
---|
Enregistrement modifié | 2020-04-16 |
---|