DOI | Trouver le DOI : https://doi.org/10.1364/AO.25.000554 |
---|
Auteur | Rechercher : Boivin, L. P.1; Rechercher : McNeely, F. T.1 |
---|
Affiliation | - Conseil national de recherches du Canada
|
---|
Format | Texte, Article |
---|
Résumé | New electrically calibrated absolute radiometers (ESRs) were developed having fast response and high responsivity; these ESRs are specially useful for measurement automation. They are fabricated using thin film techniques and have a responsivity of ∼0.3 V/W, a time constant between 2.5 and 3.0 s and a noise equivalent power of around 20 nW. Only a correction for the goldblack absorber reflectance is required, and the overall accuracy of the radiometers is ±0.1%. One of the main applications of absolute radiometers is to the accurate spectral calibration of secondary radiometers. The automation of this calibration procedure is discussed for the visible and near-IR spectral region. It is shown that the use of automation in conjunction with the new ESRs greatly simplifies calibrations and reduces turnaround time while allowing calibration accuracies of close to ±0.1%. |
---|
Date de publication | 1986-02-15 |
---|
Maison d’édition | OSA Publishing |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Numéro du CNRC | NRC-INMS-434 |
---|
Numéro NPARC | 8898104 |
---|