Téléchargement | - Voir le manuscrit accepté : Characterization of a 10V Programmable Josephson Voltage Standard and Comparison with a Hysteretic Josephson Standard (PDF, 1.4 Mio)
|
---|
Auteur | Rechercher : Granger, G.1; Rechercher : Gertsvolf, M.1; Rechercher : Wood, B.M.1 |
---|
Affiliation | - Conseil national de recherches du Canada. Science des mesures et étalons
|
---|
Format | Texte, Article |
---|
Conférence | The 2014 Conference on Precision Electromagnetic Measurements, August 24-29, 2014 , Rio de Janeiro |
---|
Sujet | DC voltage; DC voltage calibration; Josephson junctions; programmable Josephson voltage standard; hysteretic Josephson voltage standard; Josephson system comparison; volt age measurements |
---|
Résumé | A 10 V programmable Josephson voltage standard system has been installed at the National Research Council Canada. The current margins for the first Josephson step at either current polarity are greater than 1 mA. Measurements are made to establish the frequency offset and leakage resistance components of the uncertainty budget. Measurements are made to establish the frequency offset and leakage resistance components of the uncertainty budget... |
---|
Date de publication | 2014 |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Numéro NPARC | 21274929 |
---|
Exporter la notice | Exporter en format RIS |
---|