Téléchargement | - Voir le manuscrit accepté : The inverse-square gamma-irradiation anomaly of the nuclear enterprise 2575 large-volume ionisation chamber (PDF, 173 Kio)
|
---|
DOI | Trouver le DOI : https://doi.org/10.1093/rpd/ncu306 |
---|
Auteur | Rechercher : Bielajew, Alex F.; Rechercher : Tessier, Frederic1; Rechercher : Gamal, Islam El1 |
---|
Affiliation | - Conseil national de recherches du Canada. Science des mesures et étalons
|
---|
Format | Texte, Article |
---|
Résumé | The Nuclear Enterprises (Model 2575) 600 cc ionisation chamber is examined to discover the cause of its anomalous behaviour in inverse-square stability measurements. Measurements and Monte Carlo calculations are employed to isolate the cause of the discrepancy. It is found that most of the effect is due to the long photon attenuation pathlengths in the long side wall of the instrument. A phenomenological procedure, based on measurements, is proposed to correct for the anomaly. The procedure results in inverse-square stability to within 0.5% over a range of 1-7 meters. |
---|
Date de publication | 2014-10-13 |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Numéro NPARC | 21274933 |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | 02126d7b-2628-4658-938b-ff44a94e578b |
---|
Enregistrement créé | 2015-05-01 |
---|
Enregistrement modifié | 2020-06-04 |
---|