DOI | Trouver le DOI : https://doi.org/10.1109/CPEM49742.2020.9191750 |
---|
Auteur | Rechercher : Granger, G.1; Rechercher : Kurten Ihlenfeld, Waldemar G.1 |
---|
Affiliation | - Conseil national de recherches du Canada. Centre de recherche en métrologie
|
---|
Format | Texte, Article |
---|
Conférence | 2020 Conference on Precision Electromagnetic Measurements (CPEM 2020), August 24-28, 2020, Denver (Aurora), Colorado |
---|
Description physique | 2 p. |
---|
Sujet | AC-DC difference; AC voltage; digital sampling; Josephson Arbitrary Waveform Synthesizer; Josephson voltage standard; measurement uncertainty |
---|
Résumé | Using a Josephson Arbitrary Waveform Synthesizer (JAWS), we measure the AC-DC difference of a sampling digital voltmeter (SDVM) through direct sampling and subsampling. We also perform direct and differential sampling measurements of the amplitude of an AC voltage source, built at the National Research Council Canada (NRC), against the JAWS. Such calibrations of the SDVM and NRC voltage source could help maintain AC voltage traceability in between JAWS operations. |
---|
Date de publication | 2020-09-10 |
---|
Maison d’édition | IEEE |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Oui |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | 146e7c4c-4717-438f-bfcc-9e845f497dee |
---|
Enregistrement créé | 2020-11-16 |
---|
Enregistrement modifié | 2022-07-28 |
---|