Téléchargement | - Voir le manuscrit accepté : Investigation of systematic errors of an AC Jospehson Voltage Standard (PDF, 1.2 Mio)
|
---|
DOI | Trouver le DOI : https://doi.org/10.1109/CPEM.2012.6250646 |
---|
Auteur | Rechercher : Filipski, Piotr S.1 |
---|
Affiliation | - Conseil national de recherches du Canada. Institut des étalons nationaux de mesure du CNRC
|
---|
Format | Texte, Article |
---|
Conférence | 2012 Conference on Precision Electromagnetic Measurements (CPEM 2012), 1-6 July 2012, Washington, DC, USA |
---|
Sujet | Josephson effect; calibration; digital-analog conversion; superconducting arrays; voltage measurement; signal synthesis |
---|
Résumé | This paper presents results of experiments quantifying two systematic errors associated with the calibrations by use of an AC Josephson Voltage Standard: the change in the voltage leads error with the change in the level of probe immersion in the liquid helium, and the effect of the internal inductance of a Josephson array. The measurements were performed at frequencies 100 kHz and above, however, they can be used in evaluating an error budget at lower frequencies. |
---|
Date de publication | 2012-07-30 |
---|
Dans | |
---|
Langue | anglais |
---|
Publications évaluées par des pairs | Non |
---|
Numéro NPARC | 20955690 |
---|
Exporter la notice | Exporter en format RIS |
---|
Signaler une correction | Signaler une correction (s'ouvre dans un nouvel onglet) |
---|
Identificateur de l’enregistrement | bea7f81e-79a1-48cf-a708-db1960e23d61 |
---|
Enregistrement créé | 2012-11-14 |
---|
Enregistrement modifié | 2020-04-21 |
---|